1. Home
  2.  » Facilities

FACILITIES

Hyperbaric Pressure Chamber

The pressure chamber at soaked technologies is rated to 1200MSW. Each test comes with an automatically generated test report.

We have 3 penetrators available, 2x ½ inch BSP and 1x 7/16th UNF. As a standard we have one 7 pin bulkhead installed, we can populate penetrators electrically and hydraulically per client request.

Key Features

  • 1200 meter depth rated
  • Pressure data logger
  • Automatically generated test-report
  • 7/16 UNF and ½” BSP penetrators available for electronic/hydraulic connections
Specifications

Internal diameter

455mm

Internal length

1200mm

Pressure rating

120bar

Pressure media

Liquids

Pressure sensor resolution

0.2 bar

Pressure sensor accuracy

+-0.5%

SM4/SM7 TESTER

The Soaked Technologies SM4-7 tester is a next generation service tool for Seaeye SM4 and SM7 thrusters. It is capable of supplying the thruster with 36V DC as a safe working voltage for “on bench” diagnostics and trouble shooting.

Additionally the unit can run fully auto mated test cycles monitoring voltages, current and line isolation as well as measuring bollard thrust. It can also be used for endurance tests to prove thruster modifications.

Key Features

  • Large 5 cubic meter fresh water test tank
  • 100KG bollard thrust measurement
  • 550MΩ line isolation measurement
  • 2.8KW forced-air cooled meanwell power supplies
  • Automated test cycle
  • DC safe working voltage supply
  • Yearly calibration of A/D converters
  • Fully automatic comprehensive reporting
Specifications

PWM speed signal

12V, 250mA, 490Hz, 0-100%

Directional signal

+-24VDC +-2% 250mA

DC safe working voltage supply

36VDC +- 1%

SM4 power voltage supply

250VDC +-2%

SM7 power voltage supply

500 VDC +-2%

DC current supply

5.6A

Voltage and current measurement accuracy

+-1%

Earth resistance measurement

2KΩ to 0.1Ω +-2%

Directional line isolation measurement

200KΩ to 200MΩ +-5%

SM4 DC line isolation measurement

500KΩ to 250MΩ +- 5%

SM7 DC line isolation measurement

700KΩ to 550MΩ +-5%

Bollard thrust measurement

0-100 KGf +-5%